5.6 Status Output
The status output is an option that reports Analyzer conditions via contact closures on the rear
panel. The closures are available on a 50 pin connector on the rear panel. The contacts are NPN
transistors which can pass 50 ma of direct current. The pin assignments are listed in Table 5-8
below.
Table 5-8: Status Output Pin Assignments
Output #
Pin #
Definition
Condition
1
1,2
ZERO CAL
CLOSED IN ZERO CAL
2
3,4
SPAN CAL
CLOSED IN HIGH SPAN CALIBRATION
3
5,6
FLOW ALARM
CLOSED IF FLOW WARNING
4
7,8
TEMP ALARM
CLOSED IF ANY TEMP WARNING
5
9,10
DIAG MODE
CLOSED IN DIAG MODE
6
11,12
POWER OK
CLOSED IF SYSTEM POWER OK
7
13,14
CONC ALARM
CLOSED IF CONCENTRATION WARNING
8
27,28
LOW SPAN CAL
CLOSED IN LOW SPAN CALIBRATION
9
SPARE
10
19,20
AUTORANGE - HI
CLOSED IF IN HIGH RANGE
11
21,22
SYSTEM OK
CLOSED IF NO FAULTS PRESENT
12
23,24
RX CELL PRESS
CLOSED IF ABS PRES > 15" HG
The Status Board schematic can be found in the Appendix Drawing 01087.
5.7 RS-232 Interface
The RS-232 communications protocol allows the instrument to be connected to a wide variety
of computer based equipment. The interface provides two basic functions in the M9110AH.
1. First is a comprehensive command interface for operating and diagnosing the analyzer. The
command set can be used interactively or operated via stored program. This interface has,
in fact, more capabilities than the front panel keyboard.
2. The interface can provide an audit trail of analyzer events. In this function the port sends
out messages about instrument events like calibration or warning messages. If these
messages are captured on a printer or remote computer, they provide a continuous audit
trail of the analyzers operation and status.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...