3 SPECIFICATIONS, WARRANTY
3.1 Specifications
Operating Modes
NO/
NO
x
switching mode, NO only mode,
NO
x
only mode
Ranges
In 1 ppm increments from 5 ppm to 5,000 ppm
Single range, independent ranges or autoranging
Noise at zero
0.02 ppm RMS
Noise at span
<0.2% of reading RMS above 20 ppm
Detection Limit(Note 1)
0.04 ppm RMS
Zero Drift (Note 2)
<0.2% full scale/24 hours
Zero Drift (Note 2)
<0.4% full scale/7 days
Span Drift (Note 2)
<1% FS/24 hours
Lag Time
Switching Mode
20 sec (Note 3)
NO
x
mode
4 sec (Note 3)
Response Time
Switching Mode
95% in < 40 sec (Note 3)
NO
x
mode
95% in < 10 sec (Note 3)
Sample Flow Rate
290 10 cc/min (Including bypass)
Linearity
1% of full scale
Precision
0.5% of reading
Temperature Range
5-40
o
C
Humidity
0-95% RH non-condensing
Temp Coefficient
< 0.1% per
o
C
Voltage Coefficient
< 0.1% per V
Dimensions HxWxD
7"x17"x23.6" (18 cm x 43 cm x 61 cm)
Weight, Analyzer
43 lbs (20 kg)
Weight, Pump Pack
16 lbs (7 kg)
Power, Analyzer
100 V~ 50/60 Hz, 120 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 200 watts
Power, Analyzer
4
230 V~ 50 Hz, 2.5A
Power, Ext Pump
110 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 295 watts
Power, Ext Pump
4
230 V~ 50 Hz, 2.5A
Environmental
Installation Category (Over-voltage Category) II
Pollution Degree 2
Analog Resolution
1 part in 2048 of selected voltage or current range
Recorder Output
0-100 mV, 0-1, 5, 10v, bipolar
Current Loop Option
4-20ma isolated
Status
12 Status Outputs from opto-isolator
Measurement Units
ppm, mg/m
3
1. Defined as twice the zero noise level.
2. At constant temperature and voltage.
3. Lag & response times longer for external converter option.
4. Electrical ratings for CE Mark compliance.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...