TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-15
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
Control
Description
33
PMT_TEMP
NO
PMT cold block temperature. Typically 4300 mV for 7
C.
34
DCPS_VOLTAGE
NO
DC power supply composite voltage output. Typically
2500 mV
35
RCELL_PRESS
NO
Reaction Cell Pressure in mV. Typically 1270 mV for
5" Hg-A at sea level. Is an absolute pressure so higher
values means higher absolute pressures.
36
OZONE_FLOW
NO
Ozone flowmeter voltage. Typically 3200 mV at
250 cc/min.
37
CONV_TEMP
NO
Converter temp. Typically XXXX mV at 700 C
38
HVPS_VOLTAGE
NO
HVPS programming voltage. Output of HVPS is
1000x value present. 700 mV = 700 VDC output.
39
DAC_CHAN_0
NO
Output of NO
x
channel in mV.
40
DAC_CHAN_1
NO
Output of NO channel in mV.
41
DAC_CHAN_2
NO
Output of NO
2
channel in mV.
42
DAC_CHAN_3
NO
Test Channel output. Also second NO
x
output.
43
NOX_CONC
YES
NO
x
DAC programming voltage. The following 3
signals can be set to output specific voltages to each
DAC. Use in conjunction to ANALOG OUTPUT test
to check each DAC output channel. The value keyed in
should appear on the appropriate analog output
channel. This value overrides data being written from
the analyzer. Value reverts to instrument output when
function is exited.
44
NO_CONC
YES
NO DAC programming voltage. See above for
description.
45
NO2_CONC
YES
NO
2
DAC programming voltage. See above for
description.
46
TEST_OUTPUT
NO
Displays voltage of selected analog signal
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...