Step 2 - Calibrate the instrument:
Zero/Span Calibration Procedure
Step Number
Action
Comment
1.
Input Zero gas
Allow Zero gas to enter the sample port on the rear of the
instrument.
2.
Press CAL
The M9110AH enters the calibrate mode from sample mode.
3.
Wait 10 min
Wait for reading to stabilize at the zero value. If you wait less
than 10 minutes the final zero value may drift.
4.
Press ZERO
The ZERO button will be displayed.
5.
Press ENTR
Pressing ENTR actually changes the equations and zeros the
instrument.
6.
Press EXIT, input
Span gas
M9110AH returns to the CAL menu. Now switch gas streams
to span gas.
7.
Wait 10 min
Wait for reading to stabilize at the span value.
8.
Press SPAN
The SPAN button should be displayed. If there is no SPAN
button check the Troubleshooting Section 9.2.8 for
instructions on how to proceed. In certain circumstances at
low span gas concentrations both the ZERO and SPAN
buttons will appear.
9.
Press ENTR
Pressing ENTR actually changes the equations so that the
concentration displayed is the same as the expected span
concentration you entered above, thus spanning the instrument.
10.
Press EXIT
Pressing EXIT returns the instrument to SAMPLE mode.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...