4250 Flow Meter
Section 2 Programming
2-20
To activate the internal analog output(s), return to the main
screen and press
4
,
2
,
0
.
Additional information for the internal analog output board,
including specifications for the loops, is found in Section
Additional information for the internal analog output board,
including specifications for the loops, is found in Section 4.3.2.
CAUTION
Each 4-20 mA output represents a constant drain on the flow
meter of at least 16 mA, whether activated or not. While 4-20
mA applications are generally made in installations with com-
mercial power available, Teledyne Isco suggests the following
for those who have a 4-20 mA output in a battery-powered
installation.
Use with battery powered flow meters only:
• If the battery is continuously on charge (for example with a
Solar Panel Battery Charger).
• If the battery is very large, such as a deep-cycle or marine
type battery, or an Isco 55 Ampere-hour lead-acid battery.
• Use only one 4-20 mA output.
• Keep in mind that programming choices also affect power
consumption. Use “minimum” settings on the flow meter wher-
ever possible. (See Section 1, Table 1-6.)
Even with these circumstances, you may expect significantly
shorter charge life from your battery. To determine the effect of
this extra current draw on battery life, please refer to the sec-
tion
How to Make Battery Calculations
, at the end of Section 1.
The following menus determine the behavior of the 4-20 mA
current outputs. If you select ANALOG OUTPUT (another term
for the 4-20 mA Output) and the flow meter is equipped with the
optional internal board or the 4-20 mA external accessory has
b e e n t u r n e d o n , R A N G E , S M O O T H I N G, a n d M A N UA L
CONTROL will appear:
CAUTION
If you do not have the proper hardware installed and you press
4 - 2 - 0 and the number of analog outputs is not
zero
, the
external 4-20 mA conver ter will not work properly. If this
occurs, return to the 4 - 2 - 0 option, and at the prompt enter
0
for the number of outputs activated. This will restore the exter-
nal 4-20 mA converter capability.
ANALOG CAPABLE (‘0’ TO DISABLE)
OUTPUTS 3
❚
(‘EXIT PROGRAM’ WHEN DONE
ANALOG OUTPUT
• EXTERNAL 4-20 MA • • (RANGE) • • (SMOOTHING) • •
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......