4250 Flow Meter
Section 4 Options and Accessories
4-7
Table 4-2 Multiple Analog Output Board Specifications
Operating Temperature
0 to 140° F (–18 to 60° C)
Output Accuracy
0.5% of full-scale
Resolution 0.1%
of
full-scale (0-20 mA)
Electrical Isolation
Isolated from each other and from the flow meter.
Calibration
Factory-calibrated; no further adjustments necessary.
Maximum output distance
1500 feet (457.3 m) using 18 AWG wire.
Current Range (per loop)
0 to 20 mA
Maximum Load (per loop)
750 ohms
Analog Output 1 Terminations
Pin A
(Red wire –)
Pin C
(White wire +)
Analog Output 2 Terminations
Pin D
(Brown wire –)
Pin F
(Blue wire +)
Analog Output 3 Terminations
Pin E
(Black wire –)
Pin B
(Green wire +)
Note: If you must strip the cable further back to facilitate wiring, you will expose an orange, yellow,
and purple wire. You may disregard these wires, as they are not connected in this application.
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......