4250 Flow Meter
Section 5 Maintenance and Service
5-6
spacer disk between the the sensor mounting plate and the
screw at the front of the sensor.
4. Flush the underside of the sensor with water. Do not
remove the protective disk and round gasket from the level
sensor unless you can see that the ports are blocked with
solids.
5. If the ports are clogged and do not clear with the running
water, you may have to
carefully
remove the disk and gas-
ket. Removing the disk exposes the delicate, paper-thin
metal diaphragm of the pressure transducer.
Do not
touch the diaphragm with fingers or tools.
6. Gently flush with water, without training the stream
directly into the cavity.
Forcing water or air directly
against the diaphragm can ruin the probe.
Figure 5-2 Standard AV Probe with Mounting Plate
Figure 5-3 Standard AV Probe, Protective Disk Exposed
5.3.3 Cleaning the Low
Profile AV Probe
The low profile AV sensor is cleaned in the same manner as the
standard AV sensor. However, the transducer is encased in
stainless steel. Nevertheless, do not insert anything into the vent
holes when cleaning the transducer area.
Figure 5-4 Low Profile AV Probe Without Mounting Plate
Protective
Disk
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......