4250 Flow Meter
Section 2 Programming
2-47
If you select CONDUCTIVITY for the parameter to calibrate, the
following display will appear:
MS/CM is milli-siemens per centimeter. The
siemen
is the S.I.
(
Système Internationale
) name for the unit of conductance, which
is also the reciprocal of the
ohm
. The siemen was formerly called
the
mho
(
ohm
spelled backwards), and that term may be more
familiar to some. PPT is
parts per thousand
. Select the standard
most suitable for your application.
Then:
Then:
If you select PPT for the conductivity standard:
Then:
Then:
There is no need to calibrate the YSI 600 temperature sensor, as
it is self-calibrating.
CONDUCTIVITY CALIBRATION UNITS
• MS/CM • • PPT •
CONDUCTIVITY STANDARD
X.XX
MS/CM
PLACE PROBE IN
X.XX
MS/CM
PRESS ENTER WHEN STABLE:
X.XX
MS/CM
CALIBRATING...
PLEASE WAIT...
CONDUCTIVITY STANDARD
X.XX
PPT
PLACE PROBE IN
X.XX
PPT
PRESS ENTER WHEN STABLE:
X.XX
MS/CM
CALIBRATING...
PLEASE WAIT...
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......