4250 Flow Meter
Section 2 Programming
2-18
condition (or set of conditions) that must occur before the
sampler is enabled. You enter a value (level is an example) that
must be met before the enabling occurs.
But what if this value is met and then falls away? It is possible
for a condition to vary rapidly over a narrow range. Without hys-
teresis, the flow meter will turn the sampler off and on
repeatedly, causing a condition known as
chattering
, that would
result in very erratic operation of the sampler.
With hysteresis, you can enter a value that will keep the flow
meter from responding to insignificant changes in the enabling
condition. You should select a value for hysteresis that is narrow
enough to allow the flow meter to respond to any serious change,
but broad enough to ignore minor changes that could cause chat-
tering. Press
Enter
and the following will appear:
Press
Enter
again and the following will appear:
The next three menus may or may not appear, depending on
o t h e r s e l e c t i o n s y o u m a k e i n
P r o g r a m
. T h e y c o n c e r n
alarm/enable hysteresis set points for parameter sensing—tem-
perature, pH, D.O., and the YSI 600 Sonde outputs.
If you want to set hysteresis for any of these items, you should
enable them when you work through the program section, then
re-enter the
Setup
section (Hysteresis) and they will appear.
Note that you can have temperature alone, or temperature with
either pH or D.O., but you
must
have temperature with either
pH or D.O.
Note
You cannot have pH and D.O. at the same time, and selection
of one will prevent the other from appearing on the menus
later.
The following will appear if you are measuring temperature:
The following will appear if you are measuring pH:
The following will appear if you are measuring D. O:
(or mg/l depending on units selected in Program.)
VELOCITY ENABLE/ALARM
HYSTERESIS
XX.XXX
FT/S
(or meters/second)
FLOW RATE ENABLE/ALARM
HYSTERESIS
XX.XXX
CFS
(or other units of measure)
TEMPERATURE ENABLE/ALARM
HYSTERESIS
XX.XXX
DEG F
(or C)
pH ENABLE/ALARM
HYSTERESIS
X.XXX
pH
DO ENABLE/ALARM
HYSTERESIS
X.XXX
PPM
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......