4250 Flow Meter
Section 4 Options and Accessories
4-27
Altitude is just off the screen. You can select D.O. STANDARD if
this calibration medium is available to you. Do not select ABS
(absolute) BAROMETRIC PRESSURE unless you are at sea
level or know how to correct for this value. The barometric
pressure provided from the Weather Bureau is corrected for
altitude. Select ALTITUDE.
Then:
Enter the altitude for your location. This is critical to the probe’s
accuracy. The following display will then appear:
The reading is in milligrams of oxygen per liter. If you chose PPM
in program step 1 for D.O. units, the reading would be in parts
per million.
4.12 Installation of
Parameter Probes in
Mounting Rings
The parameter probes and their carriers snap onto Isco Spring
Rings and the base section of Isco Scissors Mounting Rings. You
can also install them in other ways with custom hardware as is
appropriate for your situation. For information on the Spring
Rings and the Scissors Mounting Ring, refer to Section 3.8.
4.13 The YSI 600 Multiple
Parameter Sonde
The YSI 600 Sonde is a multi-purpose water quality mea-
surement and data collection system. It is intended for use in
research, assessment, and regulatory compliance. The YSI 600
Sonde can measure the following water qualities:
• Dissolved Oxygen
• Conductivity
• pH
• Salinity
• Total Dissolved Solids
• Temperature
ALTITUDE UNITS OF MEASURE
FT, M
ENTER ALTITUDE
ALTITUDE =
XXXXX
FEET
(or meters)
WRAP D.O. PROBE IN MOIST CLOTH
PRESS ENTER WHEN STABLE:
X.XXX
MG/L
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......