4250 Flow Meter
Section 4 Options and Accessories
4-6
4.3.2 Internal Multiple
Analog Output Board
For those needing more than one analog output, Teledyne Isco
offers the Multiple Analog Output Board, which is installed
inside the flow meter. This board provides three isolated analog
outputs. The board is compatible with the existing external 4-20
mA output box (60-1784-039). A flow meter can use either the
internal board, the external box, or both, for a maximum number
of four analog outputs.
Note
If your flow meter has both the multiple analog output board
and the external analog converter, the internal ports will be
designated Analog Outputs 1, 2, and 3. The external converter
will be designated either “External” or “Analog Output 0.”
The Multiple Analog Output Board consumes a minimum of 16
mA per output; consequently, the flow meter sho uld be
AC-powered. If you must use battery power, you should consider
the following to extend battery life:
• Use a Solar Panel Battery Charger
• Use a larger battery: either a commercial
deep-cycle/marine type, or an Isco 55 Ampere-hour
lead-acid battery.
• Order only one analog output.
• Flow meter program choices also affect power
consumption. Use “minimum” settings, if possible. (See
Section 1, Table 1-6 and Section 5.)
The outputs from the analog output board are electrically iso-
lated from the flow meter and from each other by internal
DC-DC converters. The board uses the opening for the modem
connector for its outputs. Normally, a flow meter will not need
both the analog and modem boards. If your installation does,
however, you should contact the factory.
The analog board terminates in a 6-pin male M/S connector on
the flow meter case. Flow meters built with the analog board
option are also supplied with an output cable. This cable connects
to the wiring that runs to the equipment controlled by the 4-20
mA current loop and to the 6-pin M/S connector on the flow
meter. The cable has a 6-pin female M/S connector and is pro-
vided with stripped pigtail ends for convenient wiring.
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......