Model 300E Carbon Monoxide Analyzer
Instruction Manual
Operating Instructions
P/N 04288 REV: A
79
6.6.1.
Signal I/O Diagnostic Functions
The signal I/O diagnostic mode allows access to the digital and analog I/O in the
analyzer. Some of the digital signals can be controlled through the keyboard. See
Appendix A for a complete list of the parameters available for review under this
menu.
NOTE
Any Signal I/O values set while in the Signal I/O function remain as set until EXITING
the Signal I/O menu.
The Analyzer software then regains control of these signals upon exit.
To enter the signal I/O test mode, press:
=
SAMPLE
RANGE = 50.0 PPM
CO =
< TST TST > CAL
SETUP
EXAMPLE
SETUP C.3
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP C.3
COMM VARS
DIAG
HALT
EXIT
DIAG
SIGNAL I / O
PREV NEXT
JUMP
ENTR
EXIT
DIAG I / O
Test Signals Displayed Here
PREV NEXT
JUMP
PRNT EXIT
See Appendix D for a
complete list of
available SIGNALS
DIAG I / O
JUMP TO: 5
0 5
ENTR
EXIT
To Jump
to Signal no 5:
SAMPLE_LED
DIAG I / O
SAMPLE_LED = ON
PREV NEXT
JUMP
ON
PRNT EXIT
Exit to Return
to the
Diag Menu
SETUP C.3
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
Use the NEXT & PREV
keys to move UP or
DOWN one signal type.
Use the JUMP key go directly
to a specific Signal
(see Appendix D for Jump
address numbers.
Pressing the PRNT Key will send a
formatted printout over the Serial I/O Port.
This Requires that one of the
RS-232/485 com ports be active and
connected to a peripheral device.
Press Exit to
return to the
SAMPLE Mode
display