CALIBRATION PROCEDURES
Model 300E Carbon Monoxide Analyzer
Instruction Manual
136
P/N 04288 REV: A
EXIT
returns
to the
SETUP
Menu
SETUP C.4
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP C.4
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP C.4
DURATION:30.0 MINUTES
<SET
SET>
EDIT
EXIT
Toggle keys to set
duration for each iteration
of the sequence:
Set in Decimal minutes
from 0.1 – 60.0
SETUP C.4
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP C.4 CALIBRATE: OFF
ON
ENTR
EXIT
SETUP C.4
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle key
Between Off and
ON
SETUP C.4
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Display show:
SEQ 2) ZERO–SPAN, 2:00:30
Sequence
Delta Time
MODE Delta Days
CONTINUED FROM PREVIOUS PAGE
DELTA TIME
NOTE
If at any time an illegal entry is selected (Example: Delta Days > 367) the ENTR key will
disappear from the display.