Operating Instructions
Model 300E Carbon Monoxide Analyzer
Instruction Manual
60
P/N 04288 REV: A
In order to compensate for CPU clocks which run fast or slow, there is a variable to
speed up or slow down the clock by a fixed amount every day. To change this
variable, press:
SAMPLE
RANGE = 50.0 PPM
CO =
< TST TST > CAL
SETUP
SETUP C.3
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP C.3
COMM
VARS
DIAG HALT
EXIT
Exit Returns to
main Sample
Display
SETUP C.3
0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT
JUMP
EDIT PRNT EXIT
Enter Number
of Seconds per
day the clock
gains or loses.
SETUP C.3
1 ) DYN_ZERO=OFF
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP C.3
3 ) CLOCK_ADJ=0 Sec/Day
PREV JUMP
EDIT
PRNT EXIT
SETUP C.3
CLOCK_ADJ=
0 Sec/Day
+ 0 0
ENTR
EXIT
SETUP C.3
3 ) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP
EDIT PRNT
EXIT
SETUP C.3
ENTER VARS PASS: 818
8 1 8
ENTR
EXIT
SETUP C.3
2 ) DYN_SPAN=OFF
PREV
NEXT
JUMP
EDIT PRNT EXIT