Operating Instructions
Model 300E Carbon Monoxide Analyzer
Instruction Manual
70
P/N 04288 REV: A
6.4.7.
TEST Channel Output
When activated the TEST Analog Output channel cam be used to report the real-
time value of one of several of the Test Function viewable from the
SAMPLE
mode
display.
SAMPLE
RANGE = 50.0 PPM
CO =
< TST TST > CAL
SETUP
SETUP C.3
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP C.3
COMM VARS
DIAG
HALT
EXIT
Press Exit at
Any Time to
Return to DIAG
Menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
DIAG
ANALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG
TEST CHANNEL: NONE
NEXT
ENTR
EXIT
DIAG
TEST CHAN OUTPUT
PREV
NEXT
ENTR
EXIT
Press Exit at
Any Time to
Return to
SETUP Menu
Press Exit at
Any Time to
Return to Main
SAMPLE
Display
Continue pressin
NEXT
until …
DIAG
TEST CHANNEL: CO MEASURE
PREV
NEXT
ENTR
EXIT
Press
PREV
or
NEXT
to move up
and down list of
functions
(See Table 5-7)
Press EXIT to
return to the
DIAG Menu
Press ENTR to
select the Function
on the display and
activate the Test
Channel