Operating Instructions
Model 300E Carbon Monoxide Analyzer
Instruction Manual
66
P/N 04288 REV: A
SAMPLE
RANGE = 50.0 PPM
CO =
< TST TST > CAL
SETUP
SETUP C.3
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP C.3
RANGE MODE:
SNGL
SNGL
DUAL AUTO
ENTR EXIT
EXIT
Returns
to the Main
SAMPLE
Display
SETUP C.3
RANGE MODE:
SNGL
SNGL
DUAL AUTO
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP C.3
RANGE:
50.0 Conc
0 0 0
5 0 .0
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
Set Range value
between
1.0 and 1,000 ppm
Range mode
Defaults to
SNGL
Remember, while both
A1
and
A2
are reporting the same data within the same
Measurement Range span, their respective electronic signal levels can still be
configured differently (see Section 6.5) to meet the input requirements of different
recording devices.
6.4.4.
A1 & A2 Dual Range Mode
Selecting Dual Range mode allows the
A1
and
A2
outputs to be configured with
different measurement spans.
The analyzer software calls these two ranges Low and High. The Low range setting
corresponds with the analog output labeled
A1
on the Rear Panel of the instrument.
The Higher Range Setting corresponds with the
A2
output. While the software
names these two ranges Low and High, they do not have to be configured that way.