Model 300E Carbon Monoxide Analyzer
Instruction Manual
Operating Instructions
P/N 04288 REV: A
113
6.11.3.
Editing the Data Channel Trigger Event
To edit the list of Data Parameters associated with a specific Data Channel, press:
SETUP C.3
EVENT:ATIMER
<PREV NEXT>
ENTR
EXIT
Edit Data Channel Menu
SETUP C.3
0)
CONC: ATIMER, 2, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
FROM THE DATA CHANNEL PASSWORD MENU
Exits to the Main
Data Acquisition
Menu
Press each
Key
repeatedly to cycle through the
list of available Trigger Events (see Appendix A
for a complete list).
SETUP C.3
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP C.3
EVENT:ATIMER
<SET SET>
EDIT
EXIT
ENTR
accepts the new string
and returns to the previous
menu.
EXIT
ignores the new string
and returns to the previous
menu.