CALIBRATION PROCEDURES
Model 300E Carbon Monoxide Analyzer
Instruction Manual
134
P/N 04288 REV: A
To program the Sequence:
SAMPLE
CO = XX.XX
< TST TST > CAL CALZ CZLS
SETUP
SETUP C.4
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP C.4
SEQ 1) DISABLED
NEXT
MODE
EXIT
SETUP C.4
SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP C.4
MODE: DISABLED
NEXT
ENTR EXIT
SETUP C.4
MODE: ZERO
PREV
NEXT
ENTR EXIT
SETUP C.4
MODE: ZERO–SPAN
PREV
NEXT
ENTR
EXIT
SETUP C.4
SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT
MODE
SET
EXIT
SETUP C.4
TIMER ENABLE: ON
SET>
EDIT
EXIT
SETUP C.4
STARTING DATE: 01–JAN–02
<SET SET>
EDIT
EXIT
SETUP C.4
STARTING DATE: 01–JAN–02
0
4 SEP
0
3
ENTR
EXIT
Toggle keys to set
Day, Month & Year:
Format : DD-MON-Y
Y
SETUP C.4
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT
EXIT
CONTINUE NEXT PAGE
With
STARTING TIME