Model 300E Carbon Monoxide Analyzer
Instruction Manual
Operating Instructions
P/N 04288 REV: A
73
Table 6–6: Analog Output Voltage Range Min/Max
Range Span
Minimum Output
Maximum Output
0-100 mVDC
-5 mVDC
105 mVDC
0-1 VDC
-0.05 VDC
1.05 VDC
0-5 VDC
-0.25 VDC
5.25 VDC
0-10 VDC
-0.5 VDC
10.5 VDC
The default offset for all ranges is 0 VDC.
The following DC Current output options are also available from the factory:
Table 6–7: Analog Output Current Loop Min/Max
Range Span
Minimum Output
Maximum Output
2-20 mADC
1 mADC
21 mADC
4-20 mADC
3 mADC
21 mADC
The default offset for all ranges is 0 mADC.
Pin-outs for the Analog Output connector at the rear panel of the instrument are:
Table 6–8: Analog Output Pin Assignments
PIN
Analog Output
VDC Signal
mADC Signal
1
V Out
I Out +
2
A1
Ground
I Out -
3
V Out
I Out +
4
A2
Ground
I Out -
5
V Out
I Out +
6
A3
Ground
I Out -
7
V Out
Not Available
8
A4 (Spare)
Ground Not
Available