
Trace Nitrogen in Argon Analyzer
Maintenance
Teledyne Analytical Instruments
69
"Sys Test -- 5V "
The system test on the 5V power supply has found a problem.
SysTest15V 0x0100
"Sys Test -- 15V "
The system test on the 15V power supply has found a problem.
SysTest Analog 1 0x0200
"Sys Test DAC A "
The system test on analog channel 1 (% of range) is incorrect.
SysTest Analog 2 0x0400
"Sys Test DAC B "
The system test on analog channel 2 (Range ID) is incorrect
SysTestPreamp_0_0V 0x0800
"Sys Test Preamp 0.0V"
The system test on the preamp is incorrect at the zero level.
SysTestPreamp_0_7V 0x1000
"Sys Test Preamp 0.7V"
The system test on the preamp is incorrect at the non-zero level
InvLinearErr 0x4000
"Linearization Error "
The dataset for linearization could not be processed. Most likely,
two linearization outputs have been set for the same input.
5.6 Sensor Replacement
While it is possible to replace a sensor in the field, TAI
recommends that the analyzer be shipped back to the factory for sensor
replacement and requalification. If the sensor is replaced in the field, the
customer must contact TAI Customer Service to obtain the required
input data to linearize the new sensor.
Summary of Contents for 2120XL
Page 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Page 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Page 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Page 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Page 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...