Trace Nitrogen in Argon Analyzer
Operation
Teledyne Analytical Instruments
51
the above to ENTER at
More
a second time. This will display a third
screen of system functions as shown below.
Model Out-cal Lin
Use the
◄►
arrow keys to position the blinking over the
Model
function, and press ENTER to display your specific Model information.
4.5.1.11
A
NALOG
O
UTPUT
C
ALIBRATION
(O
UT
_C
AL
)
For enhanced accuracy, this function allows the user to calibrate the
4 to 20 mA analog output to remove any offsets inherent in the PCB. A
DMM configured as a DC ammeter is needed. The DMM should be
connected across the output terminals of the 4 to 20 mA output to
monitor the output current. The range of adjustment is approximately
+/- 10% of scale (+/- 1.6 mA).
If the analyzer is equipped with a secondary 0-1 VDC output, this
function can be used to calibrate the 0 to 1 volt output as well by using a
digital voltmeter on the 0-1 volt output.
To enter the 4 to 20 mA output adjust function:
Press the
SYSTEM
key to display the first
System
menu. Use the
◄►
arrow keys to position the blinking over the
More
function, and
press ENTER. Repeat the above to ENTER at
More
a second time. This
will display a third screen of system functions as shown below.
Model Out-cal Lin
Use the
◄►
arrow keys to position the blinking over the
Out_cal
function, and press ENTER. The following screen will appear:
Use UP/DOWN arrow to
Adjust 4 ma: 500
The current value associated with the 4 mA output will be
blinking. Use the
▲
/
▼
arrow keys to adjust it up or down. When the
desired value appears, Press ENTER to accept and save the value and
the display automatically moves to the next line for adjusting the 20 mA
setting.
Use UP/DOWN arrow to
Adjust 20 ma: 500
Summary of Contents for 2120XL
Page 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Page 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Page 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Page 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Page 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...