Operation Model
2120XL
Teledyne Analytical Instruments
58
After pressing the SPAN button, the mode selection screen appears:
Select Span
Mode : Manual (Auto)
where Manual and Automatic can be toggled back and forth using the
▲
/
▼
arrow keys. Once the proper mode is displayed, press ENTER.
The following screen appears:
Span
100.0
pp
m
<ENT> to begin Span
This screen allows the user to input the known concentration of the
span gas used. Uuse the
▲
/
▼
arrow keys to enter the span gas
concentration. Press ENTER when finished.
4.5.2.3.1 Manual Span
If manual span mode was selected, the following screen appears:
39.25
pp
m
<ENT> to Finish
In the Manual Mode, the user has the opportunity to decide when
the span settling is close enough to the actual concentration to be valid.
When the value is sufficiently close to the known concentration and
sufficiently stable, press ENTER to accept the span calibration.
4.5.2.3.2 Auto Mode Span
If AUTO mode was selected from the Span Mode Selection screen
the following screen appears:
Span Settling:
0:58
The time value displayed at the right count down to zero from the
value currently held in Calholdtimer. See Section 4.5.1.9.
When the counter reaches zero, the display switches to display the
live adc change count.
Summary of Contents for 2120XL
Page 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Page 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Page 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Page 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Page 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...