Operation Model
2120XL
Teledyne Analytical Instruments
50
When
HOLD
is selected, the analog output (4-20 mA) and the
range ID contacts will freeze on their last state before entering one of the
calibration modes. When the instrument returns to the
Analyze
mode,
either by a successful or an aborted calibration, there will be a three-
minute delay before the analog outputs and the range ID contacts start
tracking again.
The concentration alarms freeze on their last state before entering
calibration regardless of selecting
HOLD
or
TRACK
. But, when
HOLD
is selected the concentration alarms will remain frozen for the time set in
the
Calholdtimer
screen after the analyzer returns to the
Analyze
mode.
See Section 4.5.1.9
Calibration Hold Timer
.
4.5.1.9
C
ALIBRATION
H
OLD
T
IMER
(C
ALHOLDTIMER
)
This function sets the time period for the analog output hold as well
as the time period for which the alarms will be suppressed during a
calibration event.
To set the time period for a hold, press the
SYSTEM
key to display
the first System menu. Use the
◄►
arrow keys to position the blinking
over the
More
function, and press ENTER. A second screen of system
function will appear. Use the
◄►
arrow keys to position the blinking
over the
Calholdtimer
function, and press ENTER to advance to the
Calholdtimer
screen.
Calbrt hold:
1 min
Sample hold:
1 min
The time intervals can be set from 0 (off) to 100 minutes using the
▲
/
▼
arrow keys. Press ENTER when the desired value is displayed.
The factory default is three minutes, but the delay time is program-
mable. This preference is stored in non-volatile memory so that it is
recovered if power is removed from the instrument.
4.5.1.10
T
HE
M
ODEL
S
CREEN
The Model screen displays the manufacturer, model, and software
version information for your analyzer.
To view the information in the Model function, press the
SYSTEM
key to display the first System menu. Use the
◄►
arrow keys to
position the blinking over the
More
function, and press ENTER. Repeat
Summary of Contents for 2120XL
Page 12: ...Model 2120XL Teledyne Analytical Instruments xii ...
Page 28: ...Operational Theory Model 2120XL Teledyne Analytical Instruments 16 ...
Page 46: ...Installation Model 2120XL Teledyne Analytical Instruments 34 ...
Page 76: ...Operation Model 2120XL Teledyne Analytical Instruments 64 ...
Page 82: ...Maintenance Model 2120XL Teledyne Analytical Instruments 70 ...