Trace Nitrogen in Argon Analyzer
Maintenance
Teledyne Analytical Instruments
67
American Fuses
European Fuses
Figure 5-2: Installing Fuses
5.4 System Self Diagnostic Test
1.
Press the System button to enter the system mode.
2.
Use the
◄►
arrow keys to move to More, and press Enter.
3.
Use the
◄►
arrow keys to move to Self-Test, and press
Enter.
The following failure codes apply:
Power:
Tests the sample 5V and 15V power supplies
0
OK
1
5
V
Failure
2
15
V
Failure
3
Both
Failed
Analog:
Set non-zero sample on analog output channel 1 (% of range)
and channel 2 (Range ID) and sample response.
0
OK
1
Analog
channel
1
(%
of Range) is incorrect
2
Analog
channel
1
(Range ID) is incorrect
3
Both
channels
are
incorrect
Preamp: A)
Tests the bypass sensor input.
B)
Sets a zero level on preamp then tests response.
C)
Sets a non-zero livel on preamp then tests response.
0
OK
1
Preamp is incorrect at zero level
2
Preamp is incorrect at non-zero level
3
Both
zero
and
non-zero are incorrect