Operator’s Manual Simple TrueArb
Application
52
•
Digital Channels:
it is possible to enable up to 8 digital line on two channels model, up to 16 on
four channels model and up to 32 on eight channels model.
If the Digital Channels number is 0, the DIG button
will be disabled. If two or more
digital lines are selected the DIG button
can be touched to enable/disable the digital
lines. Once the digital channels are enabled (
≥
2 digital channels selected) it is possible to access
to the Digital Channels Page (see the picture below) to define the digital waveform sequence as
for the analog channel page.
IMPORTANT NOTE:
enabling the digital lines will cause a decrease of resolution in the analog
output channels as shown in the following table
Number
of digital
lines
CH1
Resolution
CH2
Res.
CH3
Res.
CH4
Res.
CH5
Res.
CH6
Res.
CH7
Res.
CH8
Res.
0
16 bits
16 bits 16 bits
16 bits
16 bits
16 bits
16 bits
16 bits
2
14 bits
16 bits 16 bits
16 bits
16 bits
16 bits
16 bits
16 bits
4
12 bits
16 bits 16 bits
16 bits
16 bits
16 bits
16 bits
16 bits
6
12 bits
14 bits 16 bits
16 bits
16 bits
16 bits
16 bits
16 bits
8
12 bits
12 bits 16 bits
16 bits
16 bits
16 bits
16 bits
16 bits
10
12 bits
12 bits 14 bits
16 bits
16 bits
16 bits
16 bits
16 bits
12
12 bits
12 bits 12 bits
16 bits
16 bits
16 bits
16 bits
16 bits