Operator’s Manual Simple TrueArb
Application
46
•
Single/Burst
: when the RUN/STOP button is pressed the instrument waits for a trigger event.
When the trigger event occurs each waveform will loop as written in the entry repetition
parameter and the entire sequence will be repeated circularly many times as written in the
Burst Count[N] parameter. If you set Burst Count[N]=1 the instrument is in Single mode and the
sequence will be repeated only once.
•
Triggered Continuous
: when pressing the RUN/STOP button the instrument waits for a trigger
event. When the trigger event occurs each waveform will loop as written in the entry repetition
parameter and the entire sequence will be repeated circularly until the user will press the
RUN/STOP button.
1
∞
2
3
3
2
1
N if burst / 1 if single
Start Trigger
1
2
3
3
2
1