Operator’s Manual Simple TrueArb
Application
39
IMPORTANT NOTES
•
The maximum value of repetitions is infinite: Repetitions[N]=Infinite
•
Entry Length[N]=2048 is the length of the selected sequencer entry
•
The minimum entry length is 32 samples. The entry length granularity is :
o
1 if the entry length is > 384
o
16 if entry length is
≥
32 and
≤
384 samples
•
The Waveform length is the original length of the waveform in term of number of samples
•
The entry length can be different from the waveform length because the entry length is the
same for all the instrument channels while the waveform length can be different. In case the
entry length and the waveform length are different the original waveform will be manipulated
(resampled/cut/extended) to match the entry length
•
The maximum value of repetitions is infinite. To set the repetitions to infinite open the On-
Screen Keyboard
and press the “INF”
button.
You can touch the parameter area to open the Virtual numeric keypad, edit the parameter value and its
measure unit.