Environmental requirements
Before using this product, ensure that its operating environment is maintained within these parameters:
Temperature
Ope5 °C to +40 °C (+41 °F to 104 °F)
Non-operating -20 °C to +60 °C (-4 °F to 140 °F)
Humidity
Operating 5% to 80% relative humidity with a maximum wet
bulk temperature of 29 °C at or below +40 °C, non-condensing.
Non-operating 5% to 95% relative humidity with a maximum wet
bulb temperature of 40 °C at or below +60 °C, non- condensing.
Altitude
Operating 3,000 m (9,843 feet)
Non-operating 12,000 m (39,370 feet)
Power supply requirements
WARNING.
To reduce the risk of fire and shock, ensure that the mains supply voltage fluctuations do
not exceed 10% of the operating voltage range.
No manual voltage selection is required because the AC Adapter automatically adapts to line voltage.
Source Voltage and Frequency 100 to 240 VAC ±10% @ 45-66 Hz
Power Consumption
T3AWG3352/3252 Maximum: 100W
T3AWG3354/3254 Maximum: 130W
T3AWG3358/3258 Maximum: 150W
WARNING - Electrical Shock Hazard
Only use the power cord provided with your instrument
Cleaning
WARNING.
To avoid personal injury, power off the instrument and disconnect it from line voltage
before performing any other following procedures.
Inspect the arbitrary waveform generator as often as operating conditions require. To clean the exterior
surface, perform the following steps: