Removal and Installation Procedures
6- 24
TLA7AXX Logic Analyzer Module Service Manual
Use the following procedure and Figure 6--14 to install the EMI gaskets:
1.
Position each gasket so the gasket fingers face the outside of the module.
CAUTION.
To avoid breaking the gasket fingers, do not lift the fingers too high.
2.
Pick up each gasket at the end where the gasket finger is formed up. Then
rotate the gasket on. As you do this, lift up any fingers that bind to the
chassis or cover.
3.
Slide each gasket gently from side to side to ensure that the gasket snaps in
place.
Slide open end onto module
Rotate into place lifting any
gasket fingers that bind.
Make sure gasket snaps into
holes by pulling on gasket
1
2
3
Figure 6- 14: Rear EMI gasket replacement
4.
Reinstall the Acquisition board and the LPU board if you removed them to
install the rear EMI gaskets (see pages 6--17 and 6--22).
5.
Install the module covers
.
(see page 6--13).
Side EMI Gaskets
To remove the side EMI gaskets, remove the covers and circuit boards as
necessary to access the EMI gaskets (refer to Figure 10--1 on page 10--4 for the
gasket locations). Pop the EMI gaskets out of the cover.
To install the side EMI gaskets, pop them into place on the covers.
Installation
Summary of Contents for TLA7AA1
Page 5: ......
Page 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Page 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Page 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Page 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Page 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Page 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...