Performance Verification
TLA7AXX Logic Analyzer Module Service Manual
4- 29
Start the Software.
Complete the following steps to start the software and to run
the procedures:
1.
Exit the TLA application on the logic analyzer.
2.
Double-click the TLA7Axx Performance Verification icon on the desktop.
An application window similar to Figure 4--2 on page 4--5 appears.
3.
If you have more than one logic analyzer module in your mainframe, select
the logic analyzer module from the list in the window.
4.
Click the Verification button to start the software. A Verification dialog box
similar to Figure 4--4 on page 4--17 appears.
5.
Verify that Setup and Hold is selected under Setup and Hold Procedures.
6.
Clear all other procedures in the dialog box.
NOTE
. The software will perform all selected procedures without interruption. To
limit the test to specific procedures, clear any procedures that you do not want to
run. If you want to run all of the selected procedures, you will need to play close
attention to the connection instructions as they are displayed on the screen.
7.
Click the Next button at the bottom of the dialog box to begin the procedure.
A list of connection instructions displays for the procedure.
NOTE
. Some of the connection information may vary depending on the configu-
ration of your logic analyzer module. The connection information in the window
should track your logic analyzer module; the printed instructions in this section
may differ because they assume you are testing a 136-channel module.
8.
Verify that your connections match those of the instructions. Then click the
Next button to begin the procedure.
The software will begin the procedures and display a list of results in the
window.
9.
After the first set of procedures are done, click the Next button to display the
instructions for the next step. Move the probes on the test fixture as indicated
by the instructions.
10.
After changing the connections, click the Next button to continue the
procedures.
11.
Repeat steps 8 through 9 as indicated by the software.
Summary of Contents for TLA7AA1
Page 5: ......
Page 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Page 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Page 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Page 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Page 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Page 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...