Performance Verification
TLA7AXX Logic Analyzer Module Service Manual
4- 21
3.
Connect a logic analyzer probe from the logic analyzer C0/C1/C2/C3 input
connector (or the C3/C2/A3/A2 connector for 34- or 68-channel modules) to
the appropriate connectors on the test fixture as described in the following
steps.
4.
If you are using a P6810 probe:
a.
Connect the lead sets to the P6810 GAIN & OFFSET Data connections
at J100, J101, J110, and J111 on the test fixture.
b.
Connect the clock/qualifier leads to the SNGL square pins at J220. Make
sure that you connect the ground side or the “--” side to the square pins
with the GND label.
5.
If you are using a P6860 probe, connect the probe heads to the P6860/80
GAIN & OFFSET connections at J120 and J130 on the test fixture. If you
are testing a P6880 probe, connect the probe heads at J120, J121, J130, and
J131 on the test fixture.
NOTE
. Make sure that the alignment pin on the P6860 or P6880 probe head is
aligned with the hole on the test fixture. Tighten the probe head screws by
alternating between them until they are finger tight (no more than 1 in-lbs of
torque).
6.
Power on the test fixture, logic analyzer, and the other test equipment. Allow
all test equipment to warm up for 30 minutes before continuing the
procedure.
Start the Software.
Complete the following steps to start the software and to run
the procedures:
1.
Exit the TLA application on the logic analyzer.
2.
Double-click the TLA7Axx Performance Verification icon on the desktop.
An application window similar to Figure 4--2 on page 4--5 appears.
3.
If you have more than one logic analyzer module in your mainframe, select
the logic analyzer module from the list in the window.
4.
Click the Verification button to start the software. A Verification dialog box
similar to Figure 4--4 on page 4--17 appears.
5.
Select the probe type near the bottom of the dialog box.
6.
Verify that Threshold and Analog Gain and Offset are selected under
Probe Gain & Offset Procedures. Clear all other procedures in the
dialog box.
Summary of Contents for TLA7AA1
Page 5: ......
Page 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Page 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Page 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Page 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Page 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Page 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...