Specifications
1- 12
TLA7AXX Logic Analyzer Module Service Manual
Table 1- 8: Data handling
Characteristic
Description
Nonvolatile memory retention time (
Typical
)
The battery life is integral to the NVRAM; battery life is > 10 years.
Table 1- 9: Environmental
Characteristic
Description
Temperature
Operating
5
_
C to +40
_
C/hour maximum gradient, non-condensing
The maximum operating temperature is derated at altitudes above 305 m (1000 ft.) by
1
_
C per 305 m (1000 ft.) above 1525 m (5000 ft.) altitude
Non-operating
--20
_
C to +60
_
C with 15
_
C/hour maximum gradient, without disk media
installed in disk drives.
Humidity
Operating
(No media in floppy or CD in mainframe drive):
20% to 80% relative humidity, noncondensing, maximum wet-bulb temperature of
+29
_
C (derates relative humidity to 45% at 40
_
C)
Non-operating
(No media in floppy or CD in mainframe drive):
8% to 80% relative humidity, non-condensing, maximum wet-bulb temperature of
+29
_
C (derates relative humidity to 22% relative humidity at 50
_
C)
Altitude
Operating
To 3048 m (10,000 ft.) provide maximum ambient temperature is derated by 1
_
C per
305 m (1000 ft.) above 1525m (5000 ft.)
Non-operating
12192 m (40,000 ft.)
Summary of Contents for TLA7AA1
Page 5: ......
Page 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Page 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Page 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Page 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Page 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Page 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...