
Maintenance
6- 4
TLA7AXX Logic Analyzer Module Service Manual
Inspect the internal portions of the logic analyzer module for damage and wear
using Table 6--2 as a guide. When found, defects should be repaired immediately.
If you must replace an internal component, refer to the
Removal and Installation
Procedures
, later in this chapter, for detailed removal and replacement instruc-
tions.
CAUTION.
Do not attempt to repair any circuit boards with a soldering iron.
Most of the internal components are surface mounted devices. Using a soldering
iron can damage the surface-mounted components and internal circuit boards.
Refer the replacement of surface-mounted components to qualified service
personnel with the appropriate tools.
Table 6- 2: Internal inspection check list
Item
Inspect for
Repair action
Circuit boards
Loose, broken, or corroded
connections. Burned circuit
boards. Burned, broken, or
cracked circuit-run plating.
Remove failed circuit board and
replace with a new one.
Resistors
Burned, cracked, broken, blis-
tered condition.
Remove failed circuit board and
replace with a new one.
Capacitors
Damaged or leaking cases.
Corroded solder on leads or
terminals.
Remove failed circuit board and
replace with a new one.
Semiconductors
Damaged parts or distorted pins.
Replace circuit board if parts are
damaged.
Wiring and cables
Loose plugs or connectors.
Burned, broken, or frayed wiring.
Firmly seat connectors. Repair or
replace circuit boards with defec-
tive wires or cables.
CAUTION.
To prevent damage from electrical arcing, ensure that circuit boards
and components are dry before applying power to the logic analyzer module.
To clean the interior, perform the following steps:
1.
Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).
2.
Remove any remaining dust with a lint free cloth dampened in isopropyl
alcohol (75% solution) and rinse with warm deionized water. (A cotton-
tipped applicator is useful for cleaning in narrow spaces and on circuit
boards.)
Interior Inspection
Interior Cleaning
Procedure
Summary of Contents for TLA7AA1
Page 5: ......
Page 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Page 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Page 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Page 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Page 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Page 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...