Performance Verification
4- 20
TLA7AXX Logic Analyzer Module Service Manual
8.
After the procedure finishes, click the Next button to open the Finish dialog
box where you can save the results to a file. Enter a file name and click the
Save button. You can use the Save dialog to save the file to a folder of your
choice or use the default location.
9.
Click the Finish button to finish the procedure and return to the startup
window.
CAUTION.
Always disconnect one end of each P6041 cable before powering
down the TLACAL2 test fixture. Failure to do so can damage the test fixture.
10.
Disconnect one end of each P6041 cable before powering down the
TLACAL2 test fixture.
These procedures verify the analog gain and offset parameters of the logic
analyzer module. These procedures require you to connect the logic analyzer
probes to the test fixture; the actual connections depend on the type of probe and
the number of channels of your logic analyzer module. The procedures also
verify the Threshold specifications of the logic analyzer module; this is a
certifiable parameter.
Equipment Setups.
It is assumed that the logic analyzer is already connected to
the TLACAL2 test fixture and that the performance verification software is
installed on the hard disk of the logic analyzer. It is also assumed that the logic
analyzer module is installed in the logic analyzer mainframe. The following
procedures describe the specific connections from the logic analyzer module to
the probes and to the TLACAL2 test fixture.
CAUTION.
Always power down the TLACAL2 test fixture before connecting or
disconnecting the ribbon cables. Failure to do so can damage the test fixture, the
PIC, or the MICs.
There is no need to power down the test fixture if you are only connecting
probes, the P6041 cables, or the digital multimeter to the test fixture.
1.
Connect a P6041 cable from each of the Analog Output BNC connectors to
the corresponding connector on the front of the test fixture (if necessary,
refer to the label on the test fixture for the connection locations).
2.
Connect the multimeter leads to the DMM test points (J390) on the test
fixture. Connect the positive lead to the + connector and the negative
(ground) lead to the -- connector.
Probe Gain and
Offset Procedures
Summary of Contents for TLA7AA1
Page 5: ......
Page 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Page 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Page 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Page 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Page 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Page 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...