Maintenance
TLA7AXX Logic Analyzer Module Service Manual
6- 7
Print pads
Elastomer holder
Figure 6- 1: Cleaning the probe heads
3.
Gently wipe the edge print pads of the hybrid.
4.
Remove any remaining lint using a nitrogen air gun.
5.
Put the elastomer holder back in place.
CAUTION.
Do not touch the elastomers to avoid damaging the probe contacts.
Storing the P6860 and P6880 Probe Heads
To protect the elastomer, it is important to properly store the probe heads when
the probes are not in use. See Figure 6--2.
1.
Locate the keying pin on the probe end and align it to the keying pin hole on
the nutbar.
2.
While holding the probe end at a perpendicular angle to the nutbar,
finger--tighten both probe head screws until snug (no more than 1 in-lbs of
torque).