MEAS
THS 710 & THS 720 User Manual
3–25
High-Low Setup.
The TekScope instrument determines the 10%, 50%,
or 90% levels of the selected waveform and then uses them to
calculate the measurements. You can choose the method used to
determine these levels:
H
Histogram sets the values statistically; it finds the most common
value either above or below the midpoint (depending on whether
it is defining the high or low reference level). Since this statistical
approach ignores short-term aberrations (overshoot, ringing,
noise), histogram is the best method for measuring digital
waveforms and pulses.
H
Min-max uses the highest and lowest values of the waveform
record. This method is best for measuring waveforms that have
no large, flat portions at a common value, such as sine waves and
triangle waves.
Summary of Contents for TekScope THS 710
Page 12: ...Getting Started...
Page 13: ......
Page 24: ...Operating Basics...
Page 25: ......
Page 65: ...Application Examples 2 40 THS 710 THS 720 User Manual...
Page 66: ...Reference...
Page 67: ......
Page 69: ...Introduction to Reference 3 2 THS 710 THS 720 User Manual...
Page 131: ...VERTICAL Controls 3 64 THS 710 THS 720 User Manual...
Page 132: ...Appendices...
Page 133: ......
Page 157: ...Appendix C Accessories C 6 THS 710 THS 720 User Manual...
Page 179: ...Appendix D Performance Verification D 22 THS 710 THS 720 User Manual...
Page 181: ...Appendix E General Care and Cleaning E 2 THS 710 THS 720 User Manual...
Page 182: ...Glossary and Index...
Page 183: ......