Appendix D: Performance Verification
THS 710 & THS 720 User Manual
D–13
3. Set the time mark generator period to 10 ms.
4. Set the TekScope instrument VOLTS/DIV to 500 mV/div.
5. Set the TekScope instrument SEC/DIV to 2 ms/div.
6. Use the vertical POSITION rocker to center the test signal on
screen.
7. Press SET LEVEL TO 50%.
8. Change the TekScope instrument setup using the following steps:
SCOPE
HORIZON-
TAL MENU
Time Base
Delayed Runs
After Main
—
CLEAR
MENU
—
—
9. Set the delayed time base to 500 ms/div.
10. Change the TekScope instrument setup using the following steps:
SCOPE
HORIZON-
TAL
Time Base
Delayed Runs
After Main
Set delay time
to 10 ms
11. Set the delayed time base SEC/DIV to 500 ns/div.
12. Check that the rising edge of the marker crosses the center
horizontal graticule line within
±
4 divisions of center graticule.
NOTE
. One division of displacement from graticule center corre-
sponds to a 50 ppm time base error.
Summary of Contents for TekScope THS 710
Page 12: ...Getting Started...
Page 13: ......
Page 24: ...Operating Basics...
Page 25: ......
Page 65: ...Application Examples 2 40 THS 710 THS 720 User Manual...
Page 66: ...Reference...
Page 67: ......
Page 69: ...Introduction to Reference 3 2 THS 710 THS 720 User Manual...
Page 131: ...VERTICAL Controls 3 64 THS 710 THS 720 User Manual...
Page 132: ...Appendices...
Page 133: ......
Page 157: ...Appendix C Accessories C 6 THS 710 THS 720 User Manual...
Page 179: ...Appendix D Performance Verification D 22 THS 710 THS 720 User Manual...
Page 181: ...Appendix E General Care and Cleaning E 2 THS 710 THS 720 User Manual...
Page 182: ...Glossary and Index...
Page 183: ......