Appendix D: Performance Verification
D–6
THS 710 & THS 720 User Manual
Check DC Measurement Accuracy
1. Set the DC voltage source output level to 0 V.
2. Set up the TekScope instrument using the following steps:
SCOPE
SAVE/
RECALL
Recall Saved
Setup
Recall Factory
Setup
—
OK Recall
Factory
—
CH 1
—
VERTICAL
MENU
Probe Type
Voltage Probe
Set to 1X
ACQUIRE
Mode
Average
Set to 16
MEAS
Select
Measrmnt
Mean*
—
OK Select
Measrmnt
—
*
You may need to press Select Page to display this selection.
3. Connect the TekScope instrument to the DC voltage source as
shown below.
+
–
DC voltage
source
Ch 1
Dual banana
to bnc adapter
BNC cable