Appendix D: Performance Verification
THS 710 & THS 720 User Manual
D–15
3. Set the leveled sine wave generator frequency to 100 MHz.
4. Set the TekScope instrument VOLTS/DIV to 500 mV/div.
5. Set the TekScope instrument SEC/DIV to 10 ns/div.
6. Set the leveled sine wave generator output level to approximately
50 mV
p-p
so that the measured amplitude is approximately
500 mV. (The measured amplitude can fluctuate around 500 mV.)
7. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as
necessary and then check that triggering is stable.
8. Change the TekScope instrument setup using the following steps:
SCOPE
TRIGGER
Slope
\ (falling edge)
9. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as
necessary and then check that triggering is stable.
10. Change the TekScope instrument setup using the following steps:
SCOPE
TRIGGER
Slope
/ (rising edge)
11. Proceed to the next test to check the channel 2 edge trigger
sensitivity.
Summary of Contents for TekScope THS 710
Page 12: ...Getting Started...
Page 13: ......
Page 24: ...Operating Basics...
Page 25: ......
Page 65: ...Application Examples 2 40 THS 710 THS 720 User Manual...
Page 66: ...Reference...
Page 67: ......
Page 69: ...Introduction to Reference 3 2 THS 710 THS 720 User Manual...
Page 131: ...VERTICAL Controls 3 64 THS 710 THS 720 User Manual...
Page 132: ...Appendices...
Page 133: ......
Page 157: ...Appendix C Accessories C 6 THS 710 THS 720 User Manual...
Page 179: ...Appendix D Performance Verification D 22 THS 710 THS 720 User Manual...
Page 181: ...Appendix E General Care and Cleaning E 2 THS 710 THS 720 User Manual...
Page 182: ...Glossary and Index...
Page 183: ......