HORIZONTAL Controls
3–22
THS 710 & THS 720 User Manual
Readout.
The waveform readout shows the horizontal scale factor
below the graticule. Page 3–40 shows the location of this readout.
Time Base.
Choose the Main or Delayed time base. The delayed time
base runs at the preset delay time after the trigger event for the main
time base. Use the
+
/– rocker to set the delay time.
Delay time
Main trigger
point
Main trigger
source
Delayed
waveform record
Trigger Position.
Choose the amount of pretrigger:
H
Set to 10% places the trigger point near the beginning of the
waveform record
H
Set to 50% places the trigger point at the center of the waveform
record
H
Set to 90% places the trigger point near the end of the waveform
record
H
You can also set any amount of pretrigger (0% to 100%) with
the
+
/– rocker
Summary of Contents for TekScope THS 710
Page 12: ...Getting Started...
Page 13: ......
Page 24: ...Operating Basics...
Page 25: ......
Page 65: ...Application Examples 2 40 THS 710 THS 720 User Manual...
Page 66: ...Reference...
Page 67: ......
Page 69: ...Introduction to Reference 3 2 THS 710 THS 720 User Manual...
Page 131: ...VERTICAL Controls 3 64 THS 710 THS 720 User Manual...
Page 132: ...Appendices...
Page 133: ......
Page 157: ...Appendix C Accessories C 6 THS 710 THS 720 User Manual...
Page 179: ...Appendix D Performance Verification D 22 THS 710 THS 720 User Manual...
Page 181: ...Appendix E General Care and Cleaning E 2 THS 710 THS 720 User Manual...
Page 182: ...Glossary and Index...
Page 183: ......