Appendix D: Performance Verification
D–14
THS 710 & THS 720 User Manual
Check Channel 1 Edge Trigger Sensitivity
1. Set up the TekScope instrument using the following steps:
SCOPE
SAVE/
RECALL
Recall Saved
Setup
Recall Factory
Setup
—
OK Recall
Factory
—
ACQUIRE
Mode
Average
Set to 16
TRIGGER
Mode
Normal
—
MEAS
High-Low
Setup
Min-Max
Select
Measrmnt
Ampl*
OK Select
Measrmnt
—
*
You may need to press Select Page to display this selection.
2. Connect the TekScope instrument to the leveled sine wave
generator as shown below.
Leveled
sine wave
generator
Ch 1
Output
50
feedthrough
terminator
BNC cable