Bandwidth Consideration
Bandwidth measurement system issues include the
bandwidth of both the probe and the oscilloscope.
Bandwidth is a sine wave specification. Bandwidth specifies
the maximum frequency of a sine wave that can appear on
the oscilloscope display with a maximum of 29.3% decrease
in amplitude. To ensure a sine wave amplitude error of not
more than 3%, the bandwidth of the oscilloscope and probe
combination should range between three to five times that of
the circuit being tested.
Bandwidth and rise or fall time have an inverse relationship.
The rise time of the probe and oscilloscope combination
should be three to five times less than the rise or fall time of
the measured signal. This should ensure an error of no
more than 3% in the measured rise or fall time.