6
Using Menu Function Controls
6-22
TDS1000B and TDS2000B Series Oscilloscopes – Operator Training Kit
Using the XY Display Mode
You can display waveforms in the XY or YT formats. When
you choose the XY display format, Channel 1 is displayed
on the horizontal axis and Channel 2 is displayed on the
vertical axis. When you choose the YT format, the vertical
voltage is displayed in relation to time.
You use the XY display format to measure the phase
relationship between two or more synchronous signals.
For
example, you can measure a Quadrature Amplitude
Modulation (QAM) communication signal with an XY
constellation diagram.
The following procedure enables you to see the phase
shift between two synchronous signals by displaying
the two signals in the XY format:
1. Connect the CH1 probe tip to the
5 KHz SINE
signal on
pin 10 and the CH1 ground lead to
GND
on pin 8 of the
Training 1 signal board.
2. Connect a P2220 probe to the CH2 input. Connect the
CH2 probe tip to the
PHASE SHIFTED SINE WAVE
signal on pin 11 and the CH2 ground lead to
GND
on
pin 14 of the Training 1 signal board.