6
Using Menu Function Controls
TDS1000B and TDS2000B Series Oscilloscopes – Operator Training Kit
6-9
The following procedure enables you to detect
glitches in the signal at slow time base settings by
selecting the
Peak Detect
acquisition mode:
1. Connect the CH1 probe to the
VARIABLE AMPL WITH
GLITCH
signal on pin 15 and the CH1 probe ground
lead to
GND
on pin 14 of the Training 1 signal board.
2. On the Training 1 signal board, push the
POWER
button
until only the
Analog PWR
LED is on.
3. On the top of the front panel, push
DEFAULT SETUP
.
4. In the TRIGGER section, turn the
LEVEL
knob to set
the trigger level to
400mV
on the display.
5. Push
the
TRIGGER MENU
button.
6. Push the appropriate side-screen menu button to select
Coupling HF Reject
.
Summary of Contents for TDS1000B Series