5
Using TRIGGER Controls
TDS1000B and TDS2000B Series Oscilloscopes – Operator Training Kit
5-33
The next procedure enables you to analyze the amplitude
modulation region of a complex repeating signal by
assigning a trigger holdoff in the amplitude modulation
region of the signal.
To assign a trigger holdoff for an AM signal, follow these
steps:
1. Connect the CH1 probe tip to
AM SIGNAL
on pin 9 and
the CH1 probe ground lead to
GND
on pin 8
of the
Training 1 signal board.
2. On the Training 1 signal board, push the
POWER
button
until only the
Analog PWR
LED is on.
3. On the top of the front panel, push the
DEFAULT
SETUP
button.
4. In
the
VERTICAL
section, turn the CH1
VOLTS/DIV
knob to set CH1 to
500mV
.
5. In the HORIZONTAL section, turn the
SEC/DIV
knob to
set the displayed time base to
M 50.0us
.