Using Menu Function Controls
6-26
TDS1000B and TDS2000B Series Oscilloscopes – Operator Training Kit
To measure the vertical scale of a waveform, follow these
steps:
1. Remove
the
CH2
probe tip from pin 11 PHASE SHIFT
SINE WAVE on the Training 1 signal board.
2. On the top of the front panel, push the
DEFAULT
SETUP
menu button.
3. On the top of the front panel, push the
AUTOSET
button.
4. On the top of the front panel, push the
CURSOR
menu
button.
5. Push the appropriate side-screen menu button to select
Type Amplitude
. Notice that the Multipurpose front
panel LED is now turned on.
6. Ensure
that
Source CH1
is selected.
7. Ensure that Cursor 1 is selected. Turn the
Multipurpose
knob to position the associated cursor on
the lowest point of the waveform.