Commands Listed in Alphabetical Order
The *RST command does not change the following items:
State of the USB or GPIB interface
Calibration data that affects device speci
fi
cations
Current GPIB power supply address
Stored settings
Output queue
Service Request Enable Register settings
Standard Event Status Enable Register settings
Power-On Status Clear
fl
ag setting
front-panel LOCK state
Group
Status
Syntax
*RST
*SAV (No Query Form)
Saves the state of the power supply into a speci
fi
ed nonvolatile memory location.
Any settings that had been stored previously at the location are overwritten. You
can later use the *RCL command to restore the power supply to this saved state.
Group
Status
Syntax
*SAV <NR1>
Related Commands
Arguments
<NR1>
is an integer value in the range from 1 to 40.
Examples
*SAV 2
saves the settings in memory location 2.
[SOURce:]CURRent[:LEVel]
This command sets the current value of the power supply in units of A or mA.
PWS4205, PWS4305, PWS4323, PWS4602, and PWS4721 Power Supply Programmer Manual
2-21