Commands Listed in Alphabetical Order
Related Commands
[SOURce:]VOLTage:PROTection[:LEVel]
,
Arguments
<voltage>::=<NRf><units>
where
<NRf>
is a
fl
exible decimal that sets the maximum output voltage, ranging from 0
to the maximum nameplate voltage
<units>::={V|mV|kV}
MIN
: sets the voltage limit to the minimum value (0 V).
MAX
: sets the voltage limit to the maximum value which is slightly higher than the
maximum nameplate voltage.
DEF
: sets the voltage limit to the default value, which equals
MAX
.
Returns
<NR2>
is the voltage limit in volts.
Examples
VOLT:RANG 3.2V
VOLT:RANG?
might return
3.2
, which would be the maximum programmable
voltage in volts.
*SRE
(Service Request Enable) sets and queries the bits in the Service Request Enable
Register (SRER). Refer to the
Status and Events
chapter for more information.
Group
Status
Syntax
*SRE <NR1>
*SRE?
Related Commands
,
,
Arguments
<NR1>
is an integer value in the range from 0 to 255. The binary bits of the SRER
are set according to this value. Using an out-of-range value causes an execution
error. The power-on default for SRER is 0 if *PSC is 1. If *PSC is 0, the SRER
maintains its value through a power cycle.
2-36