Chapter 2 Using the 5071A
Reading and Managing the Event Log
22
Operating and Programming Manual
Reading and Managing the Event Log
The event log contains a history of such things as power-supply status, errors, and
warnings. A maximum of 192 entries can be recorded with each numbered entry
time-and-date stamped. When the log is full, no more entries can accumulate.
Therefore, you should periodically examine the log and clear entries.
To Check Log Status
Use the following procedure to see how many entries are in the event log.
1
Select the LOG menu, then press
Enter
.
2
Select STATUS.
The number of log entries is displayed.
To View Log Entries
Use the following procedure to read event-log entries.
1
Select the LOG menu, then press
Enter
.
2
Select BROWSE, then press
Inc
or
Dec
as needed to read the entries.
To Clear the Log
This procedure clears the entire log. Cleared entries cannot be recovered.
1
Select the LOG menu, then press
Enter
.
2
Select CLEAR, then press
Enter
to clear the entire log.
To Print the Log
Use the following procedure to print the log contents when a printer is connected to the
rear-panel RS-232C serial port (See pages 41 and 43).
1
Select the LOG menu, then press
Enter
.
2
Select PRINT, then press
Enter
.
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...