Appendix A Performance Tests
5071A Complete Performance Tests
154
Operating and Programming Manual
Specifications
Standard Unit: ± 1 10
-12
High Performance unit: ± 5 10
-13
The accuracy of the 5071A is better than 5 10
-13
(High Perf) or 1 10
-12
(standard
unit). Be sure the accuracy of the reference standard is known with sufficient
precision to make this measurement accurately.
Procedure
1
The 5071A must be on for at least 30 minutes and the green continuous operation
LED must be on.
2
Connect the HP/Agilent K34-59991A Phase Comparator OUTPUT terminals to
the strip chart recorder as shown in Figure A-7. Set the recorder for 1V full scale
and 1 inch/hour, and turn on the recorder.
3
Turn on the K34-59991A power.
4
Connect the 10 MHz reference to INPUT A and the 5071A (unit under test) 10
MHz to INPUT B as shown in Figure A-7.
5
Set K34-59991A ZERO-OPER-FULL front panel mounted toggle switch to
ZERO. Adjust the ZERO SCALE control for a zero reading on the meter. Then
adjust the recorder for a zero indication.
6
Set K34-59991A switch to FULL and adjust FULL SCALE control for a full
scale reading on the recorder.
7
Check both ZERO and FULL SCALE settings on the recorder and readjust if
necessary.
8
Set K34-59991A switch to OPER for normal operation.
9
The recorder now provides a continuous record of the phase difference between
the reference standard and the 5071A unit under test. When its output reaches full
scale (360 degrees), the K34-59991A will automatically reset to 0 (0 degrees).
10
With the recorder set as described, the phase record is 100 ns full scale (with 10
MHz inputs). The figure below shows an example of a frequency difference
measurement under these conditions.
NOTE
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...